| Overall dimensions (without control console) | 2640 mm × 1120 mm × 2060 mm |
|---|---|
| C-arm longitudinal stroke | 1790 mm |
| C-arm scanning speed, maximum | 200 mm/s |
| C-arm rotation angle | 100° |
| C-arm rotation control | Motorized |
| Height | 740 mm |
|---|---|
| Length | 2400 mm |
| Width | 740 mm |
| Type | High-frequency |
|---|---|
| Maximum power | 32 kW |
| kV range | 50 — 150 kV |
| mA range | 10 — 400 mA |
| Maximum exposure time | 8 s |
| Control | From the Operator Workstation |
| Focal spot size | 0.3/0.8 mm |
|---|---|
| Type | Rotating anode |
| kV range | 50-150 kV |
| Maximum rated charts | 11 kW/49 kW @ 0.1 s |
| Anode heat capacity | 600 kHU |
| Type | Slit |
|---|---|
| Diaphragm width, not more than | 2 mm |
| Radiation angle | -12°/+12° |
| Additional filter | 4 mm Al |
| Type | CMOS linear scintillation detector with the Gd2O2S :Tb coating |
|---|---|
| Source-image distance (SID) | 1120 mm |
| Image size : whole-body (overview) large-format standard format (flat panel) | 2000 mm x 497 mm 1000 mm х 497 mm 2000 mm x 497 mm |
| Image format: whole-body (overview) large-format | 6180 pixels x 1536 pixels 3090 pixels x 1536 pixels |
| Image pixel size | 325 μm |
| High contrast limiting resolution: whole-body (overview) format large-format standard format (flat panel) | 1.5 lp/mm 2.0 lp/mm 3.0 lp/mm |
| Contrast sensitivity | 1.2 % |
| Digital signal depth | 14 bit |
| Time required for obtaining the maximum-format overview image from the moment of switching on the high voltage till obtaining the X-ray image on the monitor screen | 60 s |